All OEC 9800 and 9900 systems go through a multistep refurbishing process. EDI takes great pride in the fact that each unit is restored to a like-new status. Pain Management, Orthopedic and Rehabilitation Services are just a few of the applications utilizing the EDI refurbished OEC 9800 C-arm. The rotating anode allows for greater penetration and resolution for the lumbar area, when performing localizations or other work involving the thicker anatomical areas. The 1K X 1K CCD technology used in the OEC 9800 image system allows for crisp, clear imaging. For whatever the purpose, an EDI refurbished OEC 9800 or 9900 will deliver the results you can depend on.
Mechanical repairs/cosmetic maintenance
- GE 9800 Plus C-Arm Description: The GE OEC 9800 Plus is the gold standard in mobile fluoroscopy applications and innovative X-ray imaging technology. The OEC 9800 Plus offers full 1k 2 high-resolution imaging on a mobile system. It features proprietary image I.Q. For uncompromised image quality.
- GE OEC 9800 Plus Manuals & User Guides User Manuals, Guides and Specifications for your GE OEC 9800 Plus Medical Equipment. Database contains 1 GE OEC 9800 Plus Manuals (available for free online viewing or downloading in PDF): Quick reference manual. GE OEC 9800 Plus Quick reference manual (4 pages).
- Built on a solid foundation of leadership and experience, thousands of OEC 9800 Plus systems are installed worldwide. When you need a familiar, hard-working system to meet your clinical needs, trust the OEC 9800 to deliver with classic power and performance.
- All defective parts and covers are repaired or replaced
- Monitor cart is disassembled and repainted or retouched as needed
- X-ray tube removed and inspected for artifacts and internal integrity
- Image intensifier removed and inspected for any imperfections and grid integrity
- Camera removed, cleaned and tested Casters removed, cleaned and or replaced
- Vertical lift assembly is tested for drift and adjusted/replaced if necessary
- All bearings are inspected, cleaned and re-lubricated, or replaced as needed
- Control panel display is tested and replaced if necessary
- All locks and brake assemblies are inspected, cleaned, and repaired or replaced as needed
- Steering is tested and adjusted as required
- Foot switch and hand switch provided on each unit A complete set of operator and service manuals are provided
Re-assembly
Oec 9800 User Manual Troubleshooting
- All components sent to C-arm reassembly area for rebuilding
- Monitor cart rebuilt, II, tube, camera, main control and power cable installed
- When C-arm is completely rebuilt mechanically, system is sent for calibration and final testing
Ge Oec 9800 User Manual
Testing/calibration
- Image intensifier and X-ray tube are tested for stability and balance
- Image intensifier is tested for resolution and gain to be within OEM specifications
- X-ray tube bearings are tested for noise and coast time
- X-ray tube filaments and stator windings are tested
- X-ray tube radiation output is verified to be within OEM specifications
- X-ray generator high voltage tested
- Maximum dose rate is set in compliance with government regulations
- CCD camera calibrated to OEM specifications
- X-ray beam is aligned for each field size
- C-arm is tested for current leakage requirements
- All monitors are aligned, resized, centered and focused
- All monitors are adjusted for brightness and linearity
- Video system is optimized for gray scale and resolution
- All power supplies are tested and calibrated or replaced as necessary
Technical Specifications
Generator
- High frequency generator 15kW full-wave
- Up to 120 KVP and 75MA for radiographic exposures
- Continuous fluro-mode up to 20 mA
- One shot frame integration (low, medium, high)
- Full power from standard 110V 15A Or 220V 8A outlet
- Patented energy buffer design X-Ray Tube
- Rotating anode X-ray tube•Focal spots: 0.3 - 0.6 mm
- 300,000 HU Anode heat capacity
Physical Specifications
- Free Space in Arc - 31”
- Depth of Arc: 26”
- Arc orbital movement: 115 Degrees
- Left/Right wig-wag scan:+/- 20 degrees
- Vertical Travel: 18” motorized•Horizontal travel: 8”
- Lateral rotation: +/- 180 degrees
C-arm Dimensions
- Length: 75.9”•Height 69.9”
- Width: 33”
Workstation Dimensions
- Width: 27”
- Height 64”
- Depth: 27”